4 Maintenance and troubleshooting
Changing the Bravo head
70 Bravo Platform User Guide
CAUTION
Dropping the head or bumping the barrels or probes will damage the head. If the
Bravo head is not properly secured in place, it could drop unexpectedly. Ensure
that the head is securely locked into the head mount.
Adjusting VWorks settings
After installing a different Bravo head, you must ensure that the Bravo device in the
VWorks software is linked to the correct profile.
To adjust the VWorks settings for a different Bravo head:
1 In the VWorks window, ensure the correct device file is open for this Bravo
device, and then ensure that the profile selected under Bravo Properties is
correct.
IMPORTANT
If this is the first time to use this Bravo head with this device, you must create a
profile for the new Bravo head.
2 On the side of the Bravo Platform, press the power switch to the on (I ) position.
3 To initialize the Bravo Platform, click Initialize selected devices in the device file.

Table of Contents

Other manuals for Agilent Technologies Bravo Platform

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Bravo Platform and is the answer not in the manual?

Agilent Technologies Bravo Platform Specifications

General IconGeneral
BrandAgilent Technologies
ModelBravo Platform
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals