SYSTem:COMMunicate:ENABle {ON|1|OFF|0}, <interface>
SYSTem:COMMunicate:ENABle? <interface>
Disables or enables the GPIB, USB, or LAN remote interface. Also disables or enables available remote services such as
Sockets, Telnet, VXI11, and the built-in Web Interface.
The GPIB interface is an optional feature. See Models and Options for details.
Parameter Typical Return
{ON|1|OFF|0}, default ONfor all interfaces 0(OFF)or1(ON)
<interface>: {GPIB|USB|LAN|SOCKets|TELNet|VXI11|WEB|USBMtp|USBHost}
Disable the USB interface:
SYST:COMM:ENAB OFF,USB
l Enabling USBMtp allows you to transfer files from the instrument to a PC via the rear-panel USB port.
l The USBHost parameter refers to the instrument's front-panel USB port.
l If you disable the LAN interface, all associated LAN services will not be started when you power on the instrument.
l SYSTem:SECurity:IMMediate enables all interfaces USBMtp.
l For LAN, SOCKets, TELNet, VXI11, or WEB on the 34460A, you must have option 34460A-LAN or option
3446LANU.
l The interface enable settings are stored in nonvolatile memory. They do not change when power has been off,
after a Factory Reset (*RST) or after an Instrument Preset (SYSTem:PRESet).
SYSTem:COMMunicate:GPIB:ADDRess <address>
SYSTem:COMMunicate:GPIB:ADDRess?
Assigns instrument's GPIB (IEEE-488) address, which is displayed at power-on. Each device on the GPIB interface
must have a unique address.
The GPIB interface is an optional feature. See Models and Options for details.
Parameter Typical Return
0to30,default22 +15
Set GPIB address to 15:
SYST:COMM:GPIB:ADDR 15
l Your computer's GPIB interface card has its own address. Avoid using this address for any instrument on the GPIB
bus.
l You must cycle power for this command to take effect.
l GPIB is enabled or disabled by SYSTem:COMMunicate:ENABle.
l The GPIB interface is an optional feature. See Models and Options for details.
l This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instrument Preset
(SYSTem:PRESet).
l This parameter is set to its default value when the instrument is shipped from the factory and after
SYSTem:SECurity:IMMediate.
Agilent Truevolt Series DMM Operating and Service Guide 239

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals