SYSTem:COMMunicate:LAN:CONTrol?
Reads the initial Control connection port number for Sockets communications. This connection is used to send and
receive commands and queries.
Parameter Typical Return
(none) 5000 (0 if the interface does not support sockets)
Return the Control connection port number:
SYST:COMM:LAN:CONT?
l Use the Control socket connection to send a Device Clear (DCL) to the instrument or to detect pending Service
Request (SRQ) events.
l On the 34460A, this requires option 34460A-LAN or option 3446LANU.
SYSTem:COMMunicate:LAN:DHCP {ON|1|OFF|0}
SYSTem:COMMunicate:LAN:DHCP?
Disables or enables instrument's use of DHCP. The acronym DHCP stands for Dynamic Host Configuration Protocol, a
protocol for assigning dynamic IP addresses to networked devices. With dynamic addressing, a device can have a
different IP address every time it connects to the network.
ON: instrument tries to obtain an IP address from a DHCP server. If a DHCP server is found, it assigns a dynamic IP
address, Subnet Mask, and Default Gateway to the instrument.
OFF or DHCP unavailable: instrument uses the static IP address, Subnet Mask, and Default Gateway during
power-on.
If you change this setting, you must send SYSTem:COMMunicate:LAN:UPDate to activate the new
setting.
Parameter Typical Return
{ON|1|OFF|0}, default ON 0(OFF)or1(ON)
Disable DHCP:
SYST:COMM:LAN:DHCP OFF
SYST:COMM:LAN:UPDate
l Most corporate LANs have a DHCP server.
l SYSTem:SECurity:IMMediate sets this parameter to its default value.
l On the 34460A, this requires option 34460A-LAN or option 3446LANU.
l If a DHCP LAN address is not assigned by a DHCP server, then an Auto-IP will be assumed after approximately 2
minutes. An Auto-IP address has the form 169.254.nnn.nnn.
l This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instrument Preset
(SYSTem:PRESet).
240 Agilent Truevolt Series DMM Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals