DC Power Supply Operation 3
Protection Functions
U3606A User’s and Service Guide 87
NOTE
The OC protection feature is always enabled by default whenever you
select constant voltage mode. You cannot disable the OC protection.
If the OC value is set to a greater value than the OCP value, the OCP
value will be adjusted to equal the OC value.
If the OC value is set to zero, this will result in the output voltage being
dropped to zero for limiting.
The OC value is not applicable for square-wave output.
The OC value is limited by the range selected. Press Shift > Range to
select an appropriate range. You can only select a range when the
output is disabled (the SBY annunciator is illuminated).