6 Calibration Procedures
Recommended Test Equipment
174 U3606A User’s and Service Guide
Recommended Test Equipment
The test equipment recommended for the adjustment procedures is listed
below. If the exact instrument is not available, substitute with calibration
standards of equivalent accuracy.
Table 6 - 1 Recommended test equipment for adjustment procedures
Application Recommended equipment Recommended accuracy requirements
Zero calibration
Shorting plug — Dual banana plug with copper wire
short between the two input terminals
-
DC voltage Calibrator Fluke 5520A < 20% of instrument 1 year specification
DC current Calibrator Fluke 5520A < 20% of instrument 1 year specification
Calibrator Fluke 5520A < 20% of instrument 1 year specification
AC current Calibrator Fluke 5520A < 20% of instrument 1 year specification
Resistance Calibrator Fluke 5520A < 20% of instrument 1 year specification
Frequency
Function/Arbitrary
waveform generator
Agilent 33250A < 20% of instrument 1 year specification
Capacitance Calibrator Fluke 5520A < 20% of instrument 1 year specification