Calibration Procedures 6 
Calibration Security
U3606A User’s and Service Guide 179
Changing the calibration security code
The security code can only be changed from the remote interface after the 
instrument has been unsecured.
Send the CALibration:SECurity:CODE <new_code> command from the 
remote interface after the instrument has been unsecured.
See the CALIbration subsystem in the U3606A Programmer’s Reference 
for more information on the CALibration:SECurity:CODE <new_code> 
command.
Resetting the security code to the factory default
If you have forgotten the correct security code, you may follow the steps 
below to change the security code back to the factory default (ATU3606A).
To unsecure the instrument without the correct security code, follow the 
steps below.
& CAL:SEC:CODE OFF, 
ATU3606A
This command unsecures the instrument for 
calibration. The default security code is 
ATU3606A. The security code can only be 
changed when the instrument is unsecured.
& CAL:SEC:CODE ABC1234
This command changes the security code to 
“ABC1234”.
& CAL:SEC:CODE ON, ABC1234
This command secures the instrument from 
accidental or unauthorized adjustments.
If you do not have a record (or have lost the record) or the security code, 
first try the factory default code — “3606”, from the front panel or 
“ATU3606A”, from the remote interface — before you use the procedure 
below.