Verification and Performance Tests 5 
Performance Verification Tests
U3606A User’s and Service Guide 159
CV source effect (line regulation)
This test measures the change in output voltage that results from a change 
in AC line voltage from the minimum value to the maximum value.
1 Turn off the instrument and connect a digital multimeter between the 
front panel   (red) and   (black) output terminals (see Figure 5-6 on 
page 144).
2 Connect the AC power line through an AC voltage source. Adjust the AC 
voltage source to provide nominal input voltage to the instrument.
3 Turn on the instrument. Press Voltage to select the constant voltage 
mode. Ensure that the S1 (30 V/1 A) range is selected (the S1 
annunciator is illuminated). Press Shift > Range again if it is not.
4 Set the output voltage to the full rated value (30 V for S1 range) and 
the current to full scale. Enable the output.
5 Connect an additional electronic load across the front panel   (red) 
and   (black) output terminals in parallel with a digital multimeter 
(see  Figure 5- 6 on page 144).
6 Operate the electronic load in constant resistance mode, and set its 
resistance to 30 Ω . Ensure that the U3606A is operating within the 
specified limit and protection values. If not, adjust the electronic load 
by increasing the resistance so that the current drops slightly until the 
U3606A is operating within the specified limit and protection values.
7 Adjust the AC voltage source to the low line voltage limit. Record the 
output voltage reading on the digital multimeter.
8 Within a few seconds after step 7, adjust the AC voltage source to the 
high line voltage limit. Record the output voltage reading on the digital 
multimeter.
9 Compare the difference between the measurement results to the 
appropriate test limits shown in Table 5- 11.
Table 5 - 11 Constant voltage source effect verification test
Input AC line
Low line  
voltage limit
High line  
voltage limit
Error from  
nominal 1 year
100 90 V
ac
110 V
ac
±0.003 V