EasyManuals Logo

Agilent Technologies U3606A User Manual

Agilent Technologies U3606A
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #144 background imageLoading...
Page #144 background image
4 System Related Operation
Utility Menu Summary
122 U3606A User’s and Service Guide
Configuring the ramp signal parameters
The following procedure shows you how to configure the ramp signal
parameters.
1 To access the utility menu, press Shift > Utility.
2 The first utility menu item (Error) will be shown in the upper
secondary display.
3 Press or until the menu itemrAMP is shown in the upper
secondary display.
4 Press Voltage or Current if you wish to configure the ramp signal
parameters for CV or CC output. (The CV or CC annunciator
illuminates respectively when selected.)
5 Select an appropriate output range by pressing Shift > Range. The ramp
signal amplitude end position is limited by the range selected. (The S1
or S2 annunciator illuminates respectively when selected.)
Table 4 - 3 Ramp signal parameters
Ramp signal Output
Items Constant Voltage Constant Current
Range S1S2S1S2
Amplitude end position 0 V to 30.000 V 0 V to 8.400 V 0 A to 1.0000 A 0 A to 3.0000 A
Number of steps 1 step to 10000 steps
Dwelling time Fastest of the instrument output capability
[1]
[1] Typically ~300 ms per step.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U3606A and is the answer not in the manual?

Agilent Technologies U3606A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU3606A
CategoryMultimeter
LanguageEnglish

Related product manuals