2 Digital Multimeter Operation  
Making Measurements
46 U3606A User’s and Service Guide
Performing capacitance measurements
The U3606A Multimeter|DC Power Supply calculates capacitance by 
charging a capacitor with a known current for a period of time, and then 
measuring the voltage.
Measuring tips:
• For measuring capacitance values greater than 10000 μF, discharge the 
capacitor first, then select a suitable range for measurement. This will 
speed up the measurement time and also ensure that the correct 
capacitance value is obtained.
• For measuring capacitance values greater than 1 mF, a delayed 
response should be expected from the front panel display. For remote 
interface operation, increase the SCPI query timeout value. (Typically > 
10000 ms.)
• For measuring small capacitance values, press Null with the test leads 
open to subtract the residual capacitance of the instrument and leads.
Disconnect circuit power and discharge all high-voltage capacitors 
before measuring capacitance to avoid damaging the U3606A or the 
device under test. To confirm that a capacitor has fully discharged, use 
the DC voltage measurement. See “Measuring DC voltage” on page 26 
for more information.