EasyManua.ls Logo

Agilent Technologies U3606A - Using the Front Panel for Adjustments; Selecting the Adjustment Mode; Entering Adjustment Values

Agilent Technologies U3606A
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Calibration Procedures
Using the Front Panel for Adjustments
184 U3606A User’s and Service Guide
Using the Front Panel for Adjustments
This section describes the process used to perform adjustments from the
front panel. Refer to the U3606A Programmer's Reference for remote
interface commands.
Selecting the adjustment mode
Before performing any adjustments, you must first unsecure the
instrument for calibration. See “Unsecuring the instrument for
calibration” on page 176 for more information. Once unsecured, “CALib” is
displayed on the lower secondary display to indicate that you are in
adjustment mode.
Entering adjustment values
In the adjustment procedures, to enter an input calibration value from the
front panel, use these keys to position the cursor on a digit:
Key Description
Press to move the cursor to the left.
Press to move the cursor to the right.

Table of Contents

Related product manuals