Calibration Procedures 6
Adjustments procedures
U3606A User’s and Service Guide 213
5 Verify the constant voltage adjustments using the “CV programming and
readback accuracy” on page 157.
Finishing the adjustments
1 Remove all shorting plugs and connections from the instrument.
2 Reset the calibration message (“Calibration Message” on page 183).
3 Record the new calibration count (“Calibration Count” on page 182).
4 Press Shift > Exit or Shift + to exit the adjustment mode.
5 The instrument will be secured and returns to normal operation.
+S + + –S
30 Ω, 50 W load
V