EasyManua.ls Logo

Agilent Technologies U3606A - Table 6-11 Constant Current Output Adjustment

Agilent Technologies U3606A
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Calibration Procedures
Adjustments procedures
206 U3606A User’s and Service Guide
An adjustment failure is indicated by a long beep, the primary
display showing “FAiL” and a calibration error number appearing in
the upper secondary display. Correct the problem and repeat this
procedure.
7 Repeat step 3 through step 6 for each output adjustment item shown in
Table 6- 11.
8 Repeat step 3 through step 7 for the S1 (30 V/1 A) range. Press Current
to select the S1 range (the S1 annunciator is illuminated).
For adjustment item “LOAd”, connect a 30 Ω, 50 W load across the front
panel (red) output terminal and the I (red) terminal. Leave the
(black) terminal and LO (black) terminal connected.
9 Verify the constant current adjustments using the “CC programming and
readback accuracy” on page 163.
Table 6 - 11 Constant current output adjustment
Range Adjustment item Output current
S2 (8 V/3 A)
OUt-L 0 A
OUt-H 3 A
30 Ω, 50 W load

Table of Contents

Related product manuals