EasyManua.ls Logo

Agilent Technologies U3606A - Page 229

Agilent Technologies U3606A
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Procedures 6
Adjustments procedures
U3606A User’s and Service Guide 207
CV output adjustment procedure — front output terminals
Review the “Test Considerations” on page 140 and “Input Connections” on
page 141 sections before beginning this procedure.
1 Turn off the instrument and connect the front panel (red) and
(black) output terminals to the V (red) and LO (black) input
terminals.
2 Turn on the instrument. Press Voltage to enter the constant voltage
output calibration. Ensure that the S2 (8 V/3 A) range is selected (the
S1 annunciator is illuminated). Press Voltage again if it is not.
3 The primary display will show the uncalibrated value and the upper
secondary display will show the reference level of the adjustment item.
4 Configure each adjustment item shown in Table 6- 12.
5 Use or to select the adjustment item.
S1 (30 V/1 A)
OUt-L 0 A
OUt-H 1 A
LOAd
[1]
1 A with additional
30 Ω, 50 W load
[1] See step 8 for information on how to connect the LOAd adjustment item.
Table 6 - 11 Constant current output adjustment (continued)
Range Adjustment item Output current

Table of Contents

Related product manuals