Getting Started 1
Product at a Glance
U3606A User’s and Service Guide 21
Table 1 - 4 Output terminal connections for source functions
Function Output terminals (+ FORCE –) Maximum output
Constant voltage output
Amplitude: 0 V to 30 V
[1]
OCP
S1: 0 A to 1.1 A
S2: 0 A to 3.3 A
OC:
S1: 0 A to 1.05 A
S2: 0 A to 3.15 A
Constant current output
Amplitude: 0 A to 3 A
[1]
OVP:
S1: 0 V to 33 V
S2: 0 V to 8.8 V
OV:
S1: 0 V to 31.5 V
S2: 0 V to 8.4 V
Ramp output
Amplitude:
CV: 0 V to 30 V
[1]
CC: 0 A to 3 A
[1]
Number of steps: 1 to 10000 steps
Scan output
Amplitude:
CV: 0 V to 31.5 V
[1]
CC: 0 A to 3.15 A
[1]
Number of steps: 1 to 100 steps
Dwelling time: 1 to 99 s
Square-wave output
Amplitude: 0 V to 30 V
[1]
Frequency: 27 predefined values
Duty cycle: 256 steps
Pulse width: 256 steps
[1] Limited by range selected, S1 (30 V/1 A ) or S2 (8 V/3 A). See “Selecting a Range” on page 97 for more information on how
to select a suitable range.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U3606A and is the answer not in the manual?

Agilent Technologies U3606A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU3606A
CategoryMultimeter
LanguageEnglish

Related product manuals