EasyManua.ls Logo

Agilent Technologies U3606A - Performing Instrument Adjustments

Agilent Technologies U3606A
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Verification and Performance Tests
Performance Verification Tests
162 U3606A User’s and Service Guide
3 Set the output voltage to the full rated value (30 V for S1 range) and
current to full scale.
4 Connect an additional electronic load across the front panel (red)
and (black) output terminals in parallel with a oscilloscope (see
Figure 5- 8 on page 145).
5 Operate the electronic load in transient operation mode between one
half of the output full rated value and the output full rated value at a
1 Hz rate with 50% duty cycle.
6 Set the oscilloscope for AC coupling, internal synchronization, and lock
on either the positive or negative load transient.
7 Adjust the oscilloscope to display transients as shown in Figure 5- 13.
Note that the pulse width (t
2
– t
1
) of the transients at 15 mV from the
base line should be no more than 300 ms for the output.
Figure 5-13 Load transient response time

Table of Contents

Related product manuals