EasyManua.ls Logo

Agilent Technologies U3606A - Page 200

Agilent Technologies U3606A
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Calibration Procedures
Calibration Security
178 U3606A User’s and Service Guide
4 Press or to increment or decrement the digit.
5 Press when you are done.
6 If the correct security code is entered, “PASS” is displayed in the
primary display for 3 seconds, after which the instrument will enter the
adjustment mode.
7 If the incorrect security code is entered, “FAiL” is displayed in the
primary display for 3 seconds, after which the calibration security code
edit mode will appear again.
To unsecure the instrument from the remote interface
Send the CALibration:SECurity:CODE <mode>, <code> command
from the remote interface.
& CAL:SEC:CODE OFF,
ATU3606A
This command unsecures the instrument for
calibration. The default security code is
ATU3606A. The security code can only be
changed when the instrument is unsecured.
& CAL:SEC:CODE ON, ATU3606A
This command secures the instrument from
accidental or unauthorized adjustments.
OUT
SBY

Table of Contents

Related product manuals