DC Power Supply Operation 3
Square-wave Output
U3606A User’s and Service Guide 89
Setting the amplitude
1 To set the square- wave amplitude, press until the amplitude value
is displayed in the lower secondary display. Ensure that the
square- wave annunciator ( ) on the front panel is flashing. (Press
again if it is not.)
2 Press or to select which digit position to be changed.
3 Press or to increase or decrease the value.
4 Press again or Shift > Save to save the changes.
Square-wave amplitude
NOTE
Pressing again will save the changes, and display the next
square-wave parameter. (While still in the edit mode.)
Pressing Shift > Save will save the changes, and exit the edit mode.
The square-wave amplitude is limited by the range selected. Press
Shift > Range to select an appropriate range. You can only select a
range when the output is disabled (the SBY annunciator is illuminated).