EasyManua.ls Logo

Agilent Technologies U3606A - Page 80

Agilent Technologies U3606A
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2 Digital Multimeter Operation
Math Operations
58 U3606A User’s and Service Guide
Editing the null value
1 Press Null again to view and edit the null value.
2 Press or to select which digit position or range to be changed.
3 Press or to increase or decrease the value.
4 Press Shift > Save to save the changes. “donE” will be briefly displayed
in the upper secondary display. The display will then return to normal.
5 Press Shift > Exit to exit the edit mode without saving.
NOTE
In resistance measurement, the U3606A will read a non-zero value
even when the two test leads are in direct contact, because of the
resistance of these leads. Use the null function to zero-adjust the
display.
In DC voltage measurement mode, the thermal effect will influence the
accuracy. Short the test leads and press Null once the displayed value
is stable to zero-adjust the display.

Table of Contents

Related product manuals