M330460-01 Rev A 3-11
In HW trigger mode, when external trigger is received, Waiting for
Trig will change to Abort button, refer to Figure 3-24. Pressing the
Abort button aborts the ramp.
Figure 3-24. Voltage Ramp-Screen (Abort)
Exit Exits the Voltage Ramp sub menu and return to Ramp Screen Top
level menu, refer to Figure 3-19.
NOTE: You cannot exit out of Voltage Ramp Screen using HOME and UP arrow, these
buttons are disabled for the Voltage Ramp Screen; Refer to Figure 3-20.
Example 1: Creating a Voltage ramp using Software Trigger mode
Set the Volt to 25V
Set the To Volt to 50V
Set the Curr to 20A
Set the Time to 10s
Select the Trigger mode as SW (software)
Click on Initialize
Click on Trigger Ramp
Observe that Trigger Ramp button will change to Abort button
Observe the voltage ramp signal using oscilloscope
Clicking on the Abort button will abort the ramp.
Clicking on the Exit button will exit the Voltage Ramp screen.
Example 2: Creating a Voltage ramp using Hardware Trigger mode
Set the Volt to 25V
Set the To Volt to 50V
Set the Curr to 20A
Set the Time to 10s
Select the Trigger mode as HW (Hardware)
Click on Initialize
Observe that Trigger Ramp button will change to Waiting for Trig.
Give an external trigger i.e. an active high pulse of 10ms on the MOLEX
connector pin-8 (TRIGGER_IN) and pin-12 (DIN_RTN) to generate the
Voltage Ramp.
Observe that Waiting for Trig will change to Abort button.
Observe the voltage ramp signal using oscilloscope