21.6.5 Non-Invasive Blood Pressure (NIBP)
Non-Invasive Blood Pressure (NIBP) Specifications
NIBP
Measurement
Mode
Manual measurement, automatic interval measurement, and continuous
measurement
Interval Time of
Automated
Measurement
Mode
1 min, 2 min, 3 min, 5 min, 10 min, 15 min, 20 min, 30 min, 1 h, 2 h, 3 h, 4 h,
8 h.
Static Pressure
Measurement
Range
0 kPa (0 mmHg)- 40.0 kPa (300 mmHg)
Static Pressure
Measurement
Accuracy
± 0.4 kPa 3 mmHg)
NIBP
Measurement
Resolution
0.1 kPa (1 mmHg)
Maximum
Average Error of
Blood Pressure
Measurement
± 0.6 kPa 5 mmHg)
Maximum
Standard
Deviation of
Blood Pressure
Measurement
± 1.06 kPa 8 mmHg)
NIBP
Measurement
Range
Typ
e
Object
Systolic pressure
AMP
Diastolic
pressure
Adults
5.3-36.0 kPa
40-270 mmHg
2.7- 30.7 kPa
20- 230 mmHg
1.3-28.0 kPa
10-210 mmHg
Children
5.3- 26.7 kPa
2.7-22.0 kPa
1.3-20.0 kPa
Neonates
5.3-18.0 kPa
2.7-14.7 kPa
1.3- 13.3 kPa
NIBP Alarm
Range
Typ
e
Object
High limit alarm
Low limit alarm
Alarm error
Adu
lts
Systolic
pressure
Low limit+ 5
mmHg)- 270
mmHg
40 mmHg-(High
limit- 5 mmHg)
± 1 mmHg
AMP
(Low limit+ 5
20 mmHg-(High