26 Defibrillator Monitor Shift Check List
Check the Defibrillator Monitor according to the items in the following table after daily use.
If the items are normal, please tick in the "Normal" column. If there is any abnormality, please
indicate the condition in the"Abnormal Condition Description" column and wait until all items are
normal before the next use.
Equipment Name:
Operating Department:
Inspecting Department:
Project
Requirements
Normal
Abnormal
Condition
Description
Appearance of
Equipment
The equipment surface is clean and
free from defects
Button, On/Off
Button
No loosening or breakage
Battery
Install the battery, the battery
indicator is on, and the battery level is
≥60 %.
AC Power Supply
Connect the AC power supply. The AC
power indicator is on
Recording Paper
The recording paper is loaded and
sufficient
Cables, Connectors
The cable is not damaged, and the
plug pin is not loose or bent.
Conductive Cable
and Pads (Test on
Use)
The conductive cable and pads are
present. The cable is not damaged,
and the plug pin is not loose or bent.
Paddles (Test on
Use)
The paddle cable is not damaged, and
the plug pin is not loose or bent. The
paddles are correctly placed in the
paddle slots.
Therapy cable and
Pads (Test on Use)
The therapy cable and pad are
complete. The therapy cable is not
damaged, the plug pin is not loose or
bent, and the electrode is not expired.
Therapy cable
Shock Test *(Test
on Use)
Power on the device, select"Manual
Defibrillation", and connect the
therapy cable and test the load.