Charge to 200 J, press the "Discharge"
button, and the system prompts that a
shock is released normally. Remove
the test load after the test is
completed.
Power on the device, select"Manual
Defibrillation", and connect the
paddles. Put the electrode into the
slot, charge it to 10 J, and press the
"Discharge" button, and a normal
shock release prompt is given.