EasyManua.ls Logo

Amoul i6 - Noninvasive Pacing Specifications; Monitoring Specifications; Basic Monitoring Specifications

Amoul i6
119 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
85
Heart rate
Mini
mum
Samp
le
Experi
mental
Sampl
e
Shock
Decisi
on
Non-sho
ck
Decision
Actual
Performa
nce
Target
Performa
nce
Actual
90 %
One-si
ded
Lower
Confid
ence
Limit
Targe
t
90 %
One-
sided
Lowe
r
Confi
denc
e
Limit
(PVC)
Cardiac Arrest
100
110
0
110
Specificity
=100 %
Specificit
y >95 %
100 %
92 %
True predictive value: 98.18 % , false positive rate: 0.44 %
21.5 Noninvasive Pacing Specifications
External Pacing Specifications
Pacer Mode
On-demand pacing, fixed pacing
Pacing Rate
40 ppm-170 ppm, accuracy ±1.5 %
Pacing Current
0mA-200mA, ±5 % or ±5mA (the bigger shall prevail)
Pacing Waveform
Unidirectional square wave pulse with a pulse width of 20
ms±5 %
Slow Down Pacing
When this function is activated, the pacing rate will be 1/4 of the
original value.
21.6 Monitoring Specifications
21.6.1 Basic Monitoring Specifications
Basic ECG Monitoring Specifications
Conform to
GB9706.25 -2005, YY1079-2008, YY0782-2010, YY1139-2013
ECG Input
3-lead wire, 5-lead wire, paddles, and multi-function pads
Lead Selection
Defibrillation electrodes: pads and paddles
3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
ECG Wave
Display Time
≥16 s