Battery Specifications (with a fully charged new battery, at an ambient temperature of 20°C)
Rechargeable lithium-ion battery
Maximum energy, charging interval
greater than 1min, the printer is
not turned on
50Ω load, frequency 80bpm, the
current is
60mA and the printer is not
turned on
ECG is in the typical operating
mode, the printer is not turned on,
and the screen brightness is at the
lowest
Charging to 100% for less than 3h when it is powered off.
When powered on, charging to 100 % is less than 4.5h;
Continuous vital sign monitoring for 20 min after the alarm
(working conditions:
NIBP measurement interval is 15 min, blood oxygen probe and
ECG cable are connected, TEMP is not connected to the cable,
CO2 does not work, the recorder does not print, other
configurations are set to factory default values), and
defibrillation can be performed at least 6 times with maximum
energy
21.3 Defibrillation Specifications
Basic Defibrillation Specifications
Manual asynchronous defibrillation, manual synchronous defibrillation,
and AED defibrillation
For biphasic truncated exponential (BTE) waveform, the waveform
parameters can be automatically compensated according to the patient's
impedance.