93
Time Interval of
Measurement
Continuous
Measurement Mode
Mainstream
Start-up Time
< 10 s
Response Time
< 120 ms
Calibration
Automated continuous self-calibration
Manual calibration is not required
Pressure Compensation
With automatic barometric pressure compensation
Relative Humidity
10- 90 % (no condensation)
Accuracy (based on an
atmospheric pressure of
1 mmHg and no CO
2
during the inhalation
phase)
± 4 mmHg (≤ 40 mmHg)
Reading of± 10 % (40 mmHg< CO
2
76 mmHg)
Reading of ± 12 % (76 mmHg
CO
2
100 mmHg)
Accuracy of Respiratory
Rate
1/min
Deviation Due to Negative Effect of Gas and Fluid
Gas or Fluid
Concentration
Deviation from measured value (the measured
CO
2
value is 38 mmHg)
Oxygen(O
2
)
100 %
-1.3 mmHg
Nitrous Oxide
(N2O)
80 %
+ 6.5 mmHg
Halothane
4 %
+ 0.6 mmHg
Enflurane
5 %
+ 1.5 mmHg
Isoflurane
5 %
+ 1.7 mmHg
Sevoflurane
6 %
+ 2.7 mmHg
Desflurane
24 %
+ 6.6 mmHg
A dry mixture containing 5 % of CO
2
(38 mmHg) at 1 kPa and the remaining is N2.
21.7 Recording Specifications
Recording Specifications
Recording Mode
High-resolution thermal dot matrix printing
Recording Channel
A maximum of 3 waveforms can be produced simultaneously
Paper Speed
12.5 mm/s, 25 mm/s and 50 mm/s, the error is± 5 %
Width of the Recording
50 mm