82
Defibrillation
Electrode Category
External defibrillation paddles, multi-function pads, and internal
defibrillation paddles
External Paddles
It supports charging, discharging and energy selection, and has a charging
completion indicator and impedance indicator.
Synchronous
Discharge Delay
Time
Less than 60 ms (From R wave peak)
Module
External Manual
Defibrillation
Internal Manual
Defibrillation
Manual defibrillation energy
selection range
1/2/3/4/5/6/7/8/9/10/15/20/
30/50/70/100/120/150/170/2
00/300/360J
1/2/3/4/5/6/7/8/9/10/15/20/
30/50J
Patient Impedance Range
25 Ω-200 Ω
15 Ω-200 Ω
Defibrillation Charging Specifications (with ambient temperature at 20)
Manual defibrillation
AED defibrillation
During
charging
From power on
to completion
of charging
From the start
of the heart
rhythm analysis
to the
completion of
charging
From power on
to completion
of charging
200J
360J
200J
360J
200J
360J
200J
360J
Using a fully charged
new battery
<5 S
<8 S
<11 S
<14 S
<16 S
<21 S
<21 S
<26 S
Use a fully charged
new battery
after 15 discharges of
360J
<6 S
<9 S
<12 S
<15 S
<17 S
<22 S
<23 S
<27 S
Powered by 90% to
100% of the rated grid
supply voltage
<5 S
<8 S
<11 S
<14 S
<16 S
<19 S
<22 S
<24 S
360J defibrillation waveform (Load impedance is 25 Ω, 50 Ω/75 Ω/100 Ω/125Ω/150 Ω and 175 Ω respectively)