Guidelines and Manufacturer's Statement - Electromagnetic Immunity
The Defibrillator Monitor is intended for use in the following specified electromagnetic
environments, which the purchaser or user should ensure that it is used in such an
environment:
Immunity Test
IEC 60601
Test Level
Coincidence
Level
Electromagnetic Environment -
Guidelines
RF conduction
GB/T 17626.6
3 V (Valid
value) 150
kHz-80 MHz
3 V (Valid
value)
Portable and mobile RF communication
equipment should not be used closer to
any portion of the Defibrillator Monitor
than the recommended separation
distance, including cables. This distance
shall be calculated by the formula
corresponding to the transmitter
frequency.
Recommended separation distance:
wherein,
According to the maximum rated
output power of the transmitter in watts
(W) provided by the transmitter
manufacturer.
d - Recommended separation distance is
in meters (m).
The field strength of the fixed RF
transmitter shall be determined by a
survey of the electromagnetic field, and
the b shall be lower than the conformity
level in each frequency range.
Interference may occur near the
equipment marked with the following
symbols:
RF radiation GB/T
17626.3
3 V/m 80
MHz-2.5
GHz
3 V/m
Note 1: at 80 MHz and 800 MHz frequencies, the formula for the higher frequency band is
used.
Note 2: these guidelines may not be appropriate for all situations where electromagnetic