71
Inspection/Maintenance Items
Weekly
After use
12 months
24 months
Recorder test
×
ECG cable test
×
Manual
defibrillation test
Charge/discharge
function
×
Disarm of energy
Synchronous
defibrillation
Pacing test
NIBP test
Pressure test
×
Leak test
CO
2
module calibration
×
TEMP calibration
NIBP overpressure protection test
×
Electrical Safety
Test
Housing leakage
current test
×
Earth leakage
current
Patient leakage
current
Patient auxiliary
current
X: no maintenance or testing is required for this item.
O: maintenance or testing is required for this item.
Caution:
After 12 months of continuous use, upgraded or overhauled, a system check by our
professional maintenance personnel is required to ensure the normal use of the device.
It is recommended that the key accessories that are easy to wear and tear, such as
electrode pads and cables, should be inspected after each use and are recommended to be
replaced every three years.