78
the gas circuit
NIBP overpressure
protection
L
A
The gas circuit may be blocked. Check
and measure again.
NIBP pump leakage
L
A
Check the cuff and the inflation tube
NIBP measurement
timeout
M
A
A malfunction occurs during the
measurement and the system cannot
perform the analysis and calculation.
Check the patient's condition, check
connections, or replace the cuff and
measure again
Treatment
Module
Treatment system
power-on self-test error
S
B
The power-on self-test of the
treatment module is abnormal. Try to
restart and replace the low-pressure
board.
Defibrillation
malfunction
S
B
Defibrillation malfunction occurs or
both defibrillation malfunction and
pacing malfunction occur. Restart the
device. If the problem persists, please
contact the service personnel
Pacing malfunction
S
B
Failure of pacing function occurs.
Restart the device. If the problem
persists, please contact the service
personnel
Disarm of energy failed
H
B
Abnormalities have occurred in the
self-discharge circuit of the treatment
board. Replace the low-pressure
board and the high-pressure board