7-27
Cycle and Control Value Programming Operator Manual P129394-176
7.5.10 Audible Signals This setup option allows the operator to adjust selected audible
signals heard at the sterilizer control. Three signals can be adjusted.
Touchpad and end of cycle signals can be adjusted to one of three
volume levels or turned off. Only the volume level of the Alarm signal
can be adjusted. The Alarm signal cannot be turned off.
1. To access this utility, press the MENU touchpad from the Status
screen (#1), at the operating end of the sterilizer. The screen
changes to show Change Machine Setup screen (#20).
2. Press CHANGE MACHINE SETUP on the Menu screen, then
AUDIBLE SIGNALS on the Setup screen (#20). The display
advances to the Audible Signals setup screen (#24).
3. Select the signal you wish to adjust by pressing the appropriate
touchpad in the upper half of the screen.
• TOUCHPAD – This is the signal sounded by the control
whenever anyone presses a touchpad.
• END OF CYCLE – This is the signal heard when a cycle is
complete.
• ALARM – This is a two-tone signal heard during abnormal
conditions.
a. Each audible signal can be adjusted for volume.
1) First press the touchpad for the selected type of signal
(e.g., ALARM, END OF CYCLE or TOUCHPAD).
2) Once the signal has been selected, press the required
volume level (e.g., LOW, MEDIUM or HIGH).
b. TOUCHPAD and END OF CYCLE provide an OFF setting.
The ALARM signal cannot be turned off.
4. Once the appropriate Audible Signals and Signal volumes have
been selected, or if these do not need to be changed, press
EXIT to return to the Change Machine Setup screen (#20).
TOUCHPAD END OF
CYCLE
ALARM
SELECT VOLUME LEVEL
OR SELECT ANOTHER AUDIBLE SIGNAL
OFF LOW
EXIT
MEDIUM HIGH
24