EasyManua.ls Logo

Amsco C Series - 9.3.3 Pressure In Chamber; 9.3.4 Waste Temperature Probe Failure

Amsco C Series
154 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9-13
Troubleshooting Operator Manual P129394-176
Screen with Operator Instructions
9.3.3 Pressure In
Chamber
Occurs if 2 psig
pressure is sensed in
the chamber.
Screen with Service Instructions
Screen with Operator Instructions
9.3.4 Waste
Temperature Probe
Failure
Occurs if waste line
temperature reading is
outside the normal
range of 32 - 293°F
(0 - 145°C)
STATUS . . ALARM! 221
PRESSURE IN CHAMBER
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
AUTOMATICALLY EXHAUST CHAMBER TO
ATMOSPHERIC PRESSURE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
SERVICE
HELP
STATUS
PRINT
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION: 222
PRESSURE IN CHAMBER
-> 2 PSIG PRESSURE SENSED IN CHAMBER
WHEN NOT IN CYCLE
CAUSES AND CORRECTIONS:
1. SOLENOID VALVE MALFUNCTION
-> REPAIR S02
2. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
3. TRANSDUCER, CP, MALFUNCTION
-> REPAIR
-> RECALIBRATE
4. MAIN CONTROL FAILURE
-> CHECK CONTROL BOARD POWER STATUS LEDs
-> REPLACE CONTROL BOARD
-> RECALIBRATE
PAPER
FEED
EXIT
STATUS . . ALARM! 215
WASTE TEMPERATURE PROBE FAILURE
CHAMBER: 00.0 F 0.0 psig
WASTE: 00.0 F
STERILIZER WILL:
PREVENT NEW CYCLE FROM BEING STARTED
UNTIL ALARM IS CLEARED
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
PRINT
SILENCE
ALARM
PAPER
FEED

Table of Contents

Related product manuals