P129394-176 Operator Manual Sterilizer Operation
Table 6-3. Factory-Set Optional Liquid Cycle and Cycle Values
The AMSCO C Series Sterilizer is equipped with the following optional factory programmed liquid sterilization cycle
and cycle values.
Cycles Sterilize Temp.
Sterilize
Time
Dry
Time
Recommended Load
(by Chamber Size)
Validation
Standard
Liquid
*
250°F
(121°C)
45 Min. N/A
16 x 16" – 15 x 1,000 mL Flasks
20 x 20"
with Loading Car –
32
x 1,000 mL Flasks
with Rack and Shelves –
28
x 1,000 mL Flasks
ST-8: 2008
* The liquid cycle is for non-patient contact use only.
Figure 6-8. Liquid Cycle In-Cycle Screens
Cycle Selection Screen
Condition Phase Sterilize Phase Exhaust Phase
NOTE: Countdown time on display is estimated; timer self-corrects estimated time at beginning
of each phase.
Press for More Cycles