9-20
P129394-176 Operator Manual Troubleshooting
Alarm Description Screen with Service Instructions
Too Long To Seal Door
(Continued)
Alarm Description Screen with Operator Instructions
9.4.4 Too Long To
Unseal Door
Occurs if door seal
pressure does not drop
below 5 psig within
allotted time.
Screen with Service Instructions
STATUS . . SERVICE INFORMATION: 228
TOO LONG TO SEAL DOOR
-> DOOR SEAL DID NOT REACH 10 PSIG
WITHIN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. SEAL PRESSURE SWITCH MALFUNCTION
-> READJUST PS1 (PS2)
-> REPAIR PS1 (PS2)
2. SEAL NOT ACTIVATING
-> CHECK SEAL STEAM
-> CHECK SEAL EXHAUST
3. SOLENOID VALVE MALFUNCTION
-> REPAIR S35 (S36)
PAPER
FEED
EXIT
CLEAR
ALARM
STATUS . . ALARM! 229
TOO LONG TO UNSEAL DOOR
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
• REMAIN IN ALARM CONDITION UNTIL
DOOR IS UNSEALED
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
3. IF LOAD MUST BE REMOVED, REFER TO
EMERGENCY DOOR OPERATION PROCEDURE
IN OPERATING MANUAL
SERVICE
HELP
STATUS
PRINT
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION: 230
TOO LONG TO UNSEAL DOOR
-> DOOR SEAL PRESSURE NOT BELOW 10 PSIG
WITHIN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. SEAL PRESSURE SWITCH MALFUNCTION
-> READJUST PS1 (PS2)
-> REPAIR PS1 (PS2)
2. SEAL NOT RETRACTING
-> CHECK SEAL STEAM
-> CHECK SEAL EXHAUST
3. SOLENOID VALVE MALFUNCTION
-> REPAIR S37 (S38)
-> REPAIR S35 (S36)
PAPER
FEED
EXIT