9-7
Troubleshooting Operator Manual P129394-176
Alarm Description Screen with Operator Instructions
9.2.7 Door Unsealed
Occurs if steam
pressure in door seal
drops below 5 psig.
Screen with Service Instructions
Screen with Operator Instructions
9.2.8 Chamber
Pressure/Temperature
Failure
Occurs if chamber
pressure or
temperature readings
are outside the normal
steam range during
sterilize phase.
STATUS . . ALARM! 207
DOOR UNSEALED
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
AUTOMATICALLY ABORT CYCLE
EXHAUST OR AIR BREAK CHAMBER TO
ATMOSPHERIC PRESSURE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
PRINT
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION: 208
DOOR UNSEALED
-> STEAM PRESSURE IN DOOR SEAL BELOW
10 PSIG
CAUSES AND CORRECTIONS:
1. SEAL PRESSURE SWITCH MALFUNCTION
-> CHECK PS1 (PS2) CONNECTIONS
-> READJUST PS1 (PS2)
-> REPAIR PS1 (PS2)
2. SEAL NOT ACTIVATED
-> CHECK SEAL STEAM
-> CHECK SEAL EXHAUST
3. SOLENOID VALVE MALFUNCTION
-> REPAIR S35 (S36)
PAPER
FEED
EXIT
CLEAR
ALARM
STATUS . . ALARM! 219
PRESSURE/TEMPERATURE FAILURE
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
AUTOMATICALLY ABORT CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
PRINT
SILENCE
ALARM
PAPER
FEED