9-14
P129394-176 Operator Manual Troubleshooting
Alarm Description Screen with Service Instructions
Waste Temperature
Probe Failure
(Continued)
Screen with Operator Instructions
9.3.5 Atmospheric
Pressure Alarm
Occurs if atmospheric
pressure is greater than
1 psi from the calibrated
atmospheric pressure.
Screen with Service Instructions
STATUS . . SERVICE INFORMATION: 216
-> RTD PROBE, WT, OUTPUT IS OUTSIDE
NORMAL RANGE
CAUSES AND CORRECTIONS:
1. LOOSE CONNECTION IN PROBE WIRING
-> REPAIR
2. PROBE FAILED
-> REPLACE
-> RECALIBRATE
3. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4. MAIN CONTROL FAILURE
-> CHECK CONTROL BOARD POWER STATUS LEDs
-> REPLACE CONTROL BOARD
PAPER
FEED
EXIT
STATUS . . ALARM! 269
ATMOSPHERIC PRESSURE ALARM
CHAMBER: 00.0 F 0.0 psig
ALTITUDE SETTING: 0 TO 1000 FEET
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
PRINT
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION: 270
ATMOSPHERIC PRESSURE ALARM
-> THE ATMOSPHERIC PRESSURE IS
GREATER THAN 1 PSI FROM THE
CALIBRATED ATMOSPHERIC PRESSURE
CAUSES AND CORRECTIONS:
1. THE STERILIZER IS AT A HIGHER
ALTITUDE THEN WHERE IT WAS
ORIGINALLY CALIBRATED (FACTORY)
-> SET ALTITUDE
2. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
3. LOOSE CONNECTION IN TRANSDUCER WIRING
-> REPAIR
4. TRANSDUCER FAILED
-> REPLACE
-> RECALIBRATE
PAPER
FEED
EXIT
CLEAR
ALARM