5-12
P129394-176 Operator Manual Component Identification
5.8 Printouts Refer to Figure 5-9
The printout reports useful information about each cycle the sterilizer
runs. This includes the load number, which is a unique identifying
code. Each load number is printed in the following format:
• Two-digit month (e.g., April = 04)
• Two-digit day (e.g., twenty eighth day = 28)
• Two-digit daily cycle count (e.g., first cycle = 01, second = 02,
etc.).
Example of a complete load number: “042801” (April 28, first
cycle).
During the cycle, status lines on the printouts show the time the line
was printed, chamber temperature and the level of vacuum or
pressure in the chamber. Each status line also begins with a letter
code. This code indicates during which cycle phase the print line
occurred, or what kind of event caused the print line to occur.
Refer to Figure 5-9 for other printout features.