7-30
P129394-176 Operator Manual Cycle and Control Value Programming
2. Press DUPLICATE PRINT touchpad on the Change Machine
Setup screen. The display advances to screen #27.
3. At screen #27, select the appropriate print option by pressing
one of the touchpads in the lower half of the display.
DUPLICATE PRINT The sterilizer provides a second
printout of the last previous cycle after the cycle completes.
NO DUPLICATE PRINT This is the normal default setting.
The sterilizer does not provide an additional printout at the
end of the cycle.
4. Once the appropriate format has been selected, or if format
does not need to be changed, press EXIT to return to the
Change Machine Setup screen (#20).
7.6 Leaving Change
Values
1. Press EXIT at screen #20 to return to Menu screen #2.
2. Press EXIT at screen #2 to return to Status screen #1.
SELECT AUTOMATIC DUPLICATE PRINT STATUS
27
DUPLICATE
PRINT
NO
DUPLICATE
PRINT
EXIT