6-30
P129394-176 Operator Manual Sterilizer Operation
6.15 Aborting Cycles It may be necessary to end a processing cycle, possibly because
the wrong cycle was selected or the sterilizer begins functioning
incorrectly. A cycle can be aborted at any time by pressing the
ABORT touch-screen pad.
1. Touch the ABORT touch-screen pad.
• The status line on the display changes to EXHAUSTING
CHAMBER, if there is pressure in the chamber.
• The sterilizer exhausts the chamber of steam.
2. Once chamber reaches 4 psig, the sterilizer removes vapor from
the chamber for one minute.
3. Once vapor removal is over, status line changes to COMPLETE.
When complete the sterilizer chamber can be unloaded
following the instructions earlier in this section of the manual.
WARNING – BURN HAZARD:
• Sterilizer, rack/shelves and
loading car will be hot after
cycle is run. Always wear
protective gloves and apron
when removing a processed
load. Protective gloves and
apron must be worn when
reloading sterilizer following
the previous operation.
• Steam may be released from
the chamber when door is
opened. Step back from the
sterilizer each time the door
is opened to minimize
contact with steam vapor.
STATUS . . CHARGE 4
TEMP . . . 261 F
PRESS . . 27.7 psig
3.TRAY, PREVAC, 270F, 4, 30
PROJECTED CYCLE COMPLETION TIME:
12:20
MINUTES SECONDS
PAPER STATUS
FEED PRINT
ABORT
STATUS . . EXHAUST 6
TEMP . . . 0 F
PRESS . . 0 psig
CYCLE ABORTED!
10:20:34 AM 10/08/10
PAPER
FEED
MENU