9-9
Troubleshooting Operator Manual P129394-176
Alarm Description Screen with Operator Instructions
9.2.10 Exhaust Rate
Too Slow
Occurs if liquid cycle
exhaust rate is too slow.
Screen with Service Instructions
Screen with Operator Instructions
9.2.11 Recorder
Deviation Alarm
Occurs if the two
temperature sensing
elements in the
chamber drain probe
read more than 1°F
(0.5°C) apart.
STATUS . . ALARM! 243
EXHAUST RATE TOO SLOW
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
AUTOMATICALLY TRY TO COMPLETE CYCLE
TRY TO EXHAUST CHAMBER ACCORDING TO OPTIMAL
COOLING RATE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
SERVICE
HELP
STATUS
PRINT
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION: 244
EXHAUST RATE TOO SLOW
-> CHAMBER EXHAUSTED SLOWER THAN
THE EXPECTED RATE
CAUSES AND CORRECTIONS:
1. SOLENOID VALVE MALFUNCTION
-> REPAIR S40
2. SOLENOID VALVE MALFUNCTION
-> REPAIR S03
3. CHAMBER STEAM TRAP MALFUNCTION
-> REPAIR
4. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
FEED
EXIT
CLEAR
ALARM
STATUS . . ALARM! 281
RECORDER DEVIATION
CHAMBER: 00.0 F 0.0 psig
RECORDER: 300.0 F
STERILIZER WILL:
AUTOMATICALLY ABORT CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
STATUS
PRINT
SILENCE
ALARM
PAPER
FEED