9-4
P129394-176 Operator Manual Troubleshooting
Alarm Description Screen with Service Instructions
Too Long In Evacuation
(Continued)
NOTE: This alarm has
two service help
screens.
Screen with Operator Instructions
9.2.4 Too Long In Air
Break
Occurs if chamber
does not air break the
vacuum to 2 inHg within
the allotted time.
STATUS . . SERVICE INFORMATION: 203
TOO LONG IN EVACUATION
-> CHAMBER DID NOT REACH REQUIRED
VACUUM LEVEL WITHIN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. WATER PRESSURE LESS THAN 30 PSIG
-> CHECK WATER SUPPLY PIPING
2. CHAMBER DRAIN STRAINER PLUGGED
-> CLEAN
3. CHECK VALVE MALFUNCTION
-> REPAIR
4. SOLENOID VALVE MALFUNCTION
-> REPAIR
5. DOOR SEAL NOT ACTIVATED
-> CHECK SEAL
-> CHECK SEAL STEAM AND EXHAUST
MORE
HELP
STATUS
PRINT
PAPER
FEED
EXIT
STATUS . . SERVICE INFORMATION: 204
TOO LONG IN EVACUATION
-> CHAMBER DID NOT REACH REQUIRED
VACUUM LEVEL WITHIN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. LEAK IN PLUMBING
-> REPAIR
-> RUN A LEAK TEST
2. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
EXIT
PAPER
FEED
ABORT
STATUS . . ALARM! 225
TOO LONG IN AIR BREAK
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
AUTOMATICALLY TRY TO COMPLETE CYCLE
EXTEND EXHAUST TIME
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
SERVICE
HELP
STATUS
PRINT
SILENCE
ALARM
PAPER
FEED