9-22
P129394-176 Operator Manual Troubleshooting
Alarm Description Screen with Service Instructions
Chamber Temperature
Probe Failure
(Continued)
Alarm Description Screen with Operator Instructions
9.4.7 Jacket
Temperature Probe
Failure
Occurs if jacket
temperature reading is
outside the normal
range.
STATUS . . SERVICE INFORMATION: 212
CHAMBER TEMPERATURE PROBE FAILURE
-> RTD PROBE, CT, OUTPUT IS OUTSIDE
NORMAL RANGE
CAUSES AND CORRECTIONS:
1. LOOSE CONNECTION IN PROBE WIRING
-> REPAIR
2. PROBE FAILED
-> REPAIR
-> RECALIBRATE
3. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4. MAIN CONTROL FAILURE
-> CHECK CONTROL BOARD POWER STATUS LEDs
-> REPLACE CONTROL BOARD
-> RECALIBRATE
PAPER
FEED
EXIT
STATUS . . ALARM! 213
JACKET TEMPERATURE PROBE FAILURE
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
• AUTOMATICALLY ABORT CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
PRINT
SILENCE
ALARM
PAPER
FEED