Sterilizer Operation Operator Manual P129394-176
6. Following successful completion of the sterilization cycle, unload
the sterilizer as follows:
a. Open chamber door.
b. Remove load from chamber.
c. Slide shelves into chamber, verifying that position does not
interfere with door operation.
d. Close chamber door.
e. Transfer load to destination.
Figure 6-3. Sterilizer Chamber Equipped with Rack and
Shelves
6.8 Prevacuum
Sterilizer Cycles
AMSCO C Series Small Steam Prevacuum Sterilizers are shipped
with the factory-set cycles. Cycle descriptions are given in S
ECTION
6.9, through SECTION 6.14. Refer to Table 1 (page ii) for factory set,
qualified cycle settings.
Important: Applicable cycles have been validated to satisfy the
requirements outlined in Table 4-1 and Table 4-2. If cycle
parameters (sterilize time, dry time, temperature) other than those
listed are required, it is the responsibility of the healthcare facility to
consult and follow the device manufacturer’s written instructions.
NOTE: Contact STERIS Corporation for information on a wide range
of education/training programs designed to meet the educational
needs of healthcare industries.
WARNING – BURN HAZARD:
Steam may be released from
the chamber when door is
opened. Step back from the
sterilizer each time the door is
opened to minimize contact
with steam vapor.
WARNING – SLIPPING
HAZARD: To prevent falls
keep floors dry by
immediately wiping up any
spilled liquids or
condensation in sterilizer
loading or unloading area.