1-3
Safety Precautions Operator Manual P129394-176
CAUTION POSSIBLE EQUIPMENT DAMAGE:
According to AAMI standards, a measured leak rate greater than 1 mm Hg/minute (1.3 mbar/min)
indicates a problem with the sterilizer. Refer the situation to a qualified service technician before using the
sterilizer further.
Gasket must be fully retracted prior to operating sterilizer door.
If 0 dry time is selected, sterilizer automatically initiates a vapor removal phase in place of drying. This
phase can still draw a vacuum to 5 inHg. Consult device manufacturer’s recommendations to ensure
devices being processed can withstand this depth of vacuum.
Lifting the chamber float switch when cleaning the chamber may cause the sterilizer control to initiate a
“Chamber Flooded” alarm. If this alarm condition occurs, the operator must turn the control power OFF
then ON to clear the alarm. The control power switch is located in the mechanical area at the side of the
sterilizer. Placing the sterilizer in standby does not clear this alarm.
Never use a wire brush, abrasives or steel wool on door and chamber assembly. Do not use cleaners
containing chloride on stainless-steel surfaces. Chloride-based cleaners will deteriorate stainless steel,
eventually leading to failure of the vessel.
Do not use cleaners containing chlorides on loading cars. Chloride-based cleaners will deteriorate the
loading car metal.
Sterilization of chloride-containing solutions (e.g., saline) can cause chamber corrosion and is not
recommended by the manufacturer. If, however, chloride-containing solutions must be processed, clean
the chamber after each use.
Allow thermostatic traps to cool down to room temperature before removing cover. Since there is nothing
to limit expansion, the bellows may rupture or fatigue if trap is opened while hot.
Actuation at less than 75% of rated pressure can allow debris to contaminate the seat and cause the
safety valve to leak. A leaking safety valve must be replaced.
Insufficient service clearance will make repairs more difficult and time-consuming.
Piping sized too small may cause water hammer, resulting in damage to the sterilizer.
After installation, it is mandatory to brace piping at the drain funnel so that it will not move vertically.
Make sure door opening is clear of any obstruction before closing the door(s).
Do not attempt to open sterilizer door during manual operation unless chamber is at 0 psig.
During manual operation, gasket must be fully retracted prior to operating sterilizer door.
Immediately wipe up saline solution spills on loading car to prevent damage to stainless steel.
WARNING STERILITY ASSURANCE HAZARD: (CONTINUED)
!
!
!
!
!
!
!
!
!
!
!
!
!
!
!
!

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Amsco C Series and is the answer not in the manual?

Amsco C Series Specifications

General IconGeneral
BrandAmsco
ModelC Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals