5-19
Component Identification Operator Manual P129394-176
DRAIN – Boiler drain opens, allowing water to drain.
NOTE: Heater elements are OFF during Drain phase.
FLUSH – Water flows through the boiler chamber for 5, 10 or 15
minutes as selected by the user (see Section 7.5.4, Steam
Generator Flush and Drain Set Up on page 7-20).
Once FLUSH is complete, the boiler chamber fills and reheats
water. Approximately 10 - 15 minutes are required to build a
head of steam suitable for running a sterilization cycle.