5-19
Component Identification Operator Manual P129394-176
DRAIN Boiler drain opens, allowing water to drain.
NOTE: Heater elements are OFF during Drain phase.
FLUSH Water flows through the boiler chamber for 5, 10 or 15
minutes as selected by the user (see Section 7.5.4, Steam
Generator Flush and Drain Set Up on page 7-20).
Once FLUSH is complete, the boiler chamber fills and reheats
water. Approximately 10 - 15 minutes are required to build a
head of steam suitable for running a sterilization cycle.
STATUS . .DRAIN 0:00 4
TEMP . . .00 F
PRESS . . .0.0 inHg
GENERATOR FLUSH
PROJECTED CYCLE COMPLETION TIME:
12:20
MINUTES SECONDS
PAPER STATUS
FEED PRINT
ABORT
STATUS . .FLUSH 0:00 4
TEMP . . .00 F
PRESS . . .00 inHg
GENERATOR FLUSH
PROJECTED CYCLE COMPLETION TIME:
05:20
MINUTES SECONDS
PAPER STATUS
FEED PRINT
ABORT