Section 7 Sample Programs
7-42
14 '
15 Call ibwrt(Ans%, "MEAS SETCOM")'
Moves to common parameter setting
screen.
16 Call ibwrt(Ans%, "DUTCTRL NONE")'
Sets DUT Control to None.
17 Call ibwrt(Ans%, "FREQBAND D8ØØMHZ")'
Sets frequency band to digital 800 MHz
band.
18 Call ibwrt(Ans%, "CHAN 1")'
Sets measurement frequency channel to
CH1.
19 Call ibwrt(Ans%, "RFLVL 1ØDBM")'
Sets TX reference level to 10 dBm.
2Ø Call ibwrt(Ans%, "MEASOBJ MSDTC")'
Sets measurement object signal to MS-
DTC.
21 Call ibwrt(Ans%, "SLTNUM 1")'
Sets measurement slot number to 1.
22 Call ibwrt(Ans%, "DVCC Ø1")'
Sets DVCC data to 01H.
23 '
24 Call ibwrt(Ans%, "MEAS SETDRX")'
Moves to Setup Digital RX parameter
screen.
25 Call ibwrt(Ans%, "BERMEASIN RFLOOP")'
Sets BER signal input to RF connector.
26 Call ibwrt(Ans%, "SACCH ØØØ")'
Sets SACCH data to 000H.
27 '
28 End Sub
By lines 11 and 12, RF signal input terminal is set.
Setting of parameter for RX measurement is made on the Setup Common parameter
screen and the Setup Digital RX Measure Parameter screen.
Lines 15 to 22 are setting of Setup Common parameter screen. Here, test signal center
frequency, TX reference level (for loop back), and measurement signal classification
are set.
Line 24 and subsequent lines are settings at the Setup Digital RX Measure Parameter
screen. The RF signal parameter is set as the measurement object (measurement sig-
nal parameter, BER input interface, etc.).