Section 2 Preparations Before Use
2-8
2.3.2 Replacing fuse
The MT8801C with standard accessories has two spare fuses (T6.3A250V). Use
these fuses to replace the blown fuses. If the fuses must be replaced, locate and rem-
edy the cause before replacing the blown fuses.
Power supply
system
Voltage range
Fuse rating plate Fuse rating Fuse name Model/Order No.
100 Vac 100 120 V T6.3 A
6.3 A, 250 V T6.3 A 250 V F0014
200 Vac 200 240 V T6.3 A
Fuse holder
F
U
S
E
F
U
S
E
AC 100120V T 6.3A
AC 200240V T 6.3A
WARNING
Prevention of electric shock
Before replacing the fuses, turn the power switch off and
remove the power cord from the power outlet. If the fuses
are replaced while power is being supplied,there is a seri-
ous risk of electric shock.
Confirmation before turning the power on
After replacing fuses, the protective grounding mentioned
above must be provided before turning the power on
again, and the proper AC power supply voltage must be
confirmed.
If the AC power supply voltage is improper, there is a risk
of the internal circuits of the MT8801C being damaged.