6-13
6.3 Performance Tests
6.3.5 Power meter (IF Level Meter) measurement accuracy
The accuracy of IF level meter of MT8801C is measured.
(1) Specifications
±0.4 dB (+40 to 0 dBm after the Power Meter Calibration execution)
±0.4 dB (+40 to 10 dBm after the Power Meter Calibration execution at 18 to
28 ˚C)
±0.7 dB (+40 to 10 dBm after the Internal OSC Calibration execution at 18 to
28 ˚C)
Linearity (10 dBm Ref.Level or larger as reference)
0 to 10 dB: ±0.1 dB
10 to 20 dB: ±0.2 dB
20 to 40 dB: ±0.5 dB
(2) Test instruments
Linearity measurement
Receiver for calibration: ML2530A or the equivalent
Intelligent RF signal generator: HP8665B or the equivalent
Power divider
Measurement accuracy (INT OSC Calibration, Power Meter Calibration)
Intelligent RF signal generator: HP8665B or the equivalent
Programmable attenuator: MN63A or the equivalent
Power meter: ML4803A or the equivalent
Power sensor: MA4601A or the equivalent
30-dB attenuator for sensitivity adjustment: MP47A or the equivalent
N-f to N-f adapter