6-7
6.3 Performance Tests
6.3.2 Frequency/waveform quality
(1) Specifications
• Frequency measurement: Measurement error: Reference ±10 Hz
• Waveform quality: Measurement range: 0.9 to 1.0
Measurement error: ±0.003
• Residual vector error: <5 %
(The above specifications are guaranteed after the Adjust Range execution.)
(2) Test instrument
• Digital modulation signal generator: MG3670B or the equivalent
(3) Setup
MT8801C
MG3670B
Digital modulation signal generator
10 MHz Buffered Output 10 MHz Reference Input
Main Input RF Output
MT8801C
Radio Communication
Analyzer
300kHz-3GHz
Input/Output
Main
Output
300kHz-3GHz
50‰
10W Max
20dBm Max
AUX
50‰