6.6 Device Status at Power-on
6.6 Device Status at Power-on
When the power is switched on:
[1] Preset value: When a power-off time (POWERON LAST) is selected, the device
is set to the status before the last power off.
Preset value: When Recall memory No. (POWERON n) is selected, the device is
set to file (number [n]) status.
[2] The Input Buffer and Output Queue are cleared.
[3] The Purser, Execution Controller, and Response For matter are initialized.
[4] The device is put into OCIS (Operation Complete Command Idle State).
[5] The device is put into OQIS (Operation Complete Query Idle State).
[6] The Standard Event Status and Standard Event Status Enable Registers are
cleared. Events can be recorded after the registers have been cleared.
For the special case of [1], when the power supply is first turned on after the device is
shipped, the initial values are set to those in the initial setting table (refer to separate
Operation Manual Vol. 1 Appendix B).