5.2 Testing CDMA Mobile station Output
5.2.11 Measuring standby output power
Standby output power is the output power in the waiting status after the completion of
the CDMA mobile station location registration.
Measurement procedure
The explanation in this section starts at the completion of the setting explained in 5.1.2
“Setting MT8801C and the CDMA mobile station”.
(1) Press the F6 (Main Func On Off) key to turn On the Main Func.
(2) Press the F1 (TX & RX Tester) key.
(3) Press the key of Next Menu to display the second page of main function keys
(F1 to F6).
(4) Press the F5 (Stop) key to release the Loopback status.
(5) Turn OFF the CDMA mobile station power.
(6) Press the F5 (Start) key.
(7) Turn ON the CDMA mobile station power.
(8) Wait until the Call Proc. status becomes the Idle (Regist) status.
(9) Press the key of Next Menu to display the first page of main function keys (F1
to F6).
(10) Press the F1 (TX Measure) key.
(11) Press the key of Next Menu to display the second page of function keys (F7
to F12).
(12) Press the F9 (Setup TX Parameter) key.
(13) Set the following item on the Setup TX Measure Parameter screen:
♦ Set PCB Pattern to be Alternate.
(14) Press the F12 (Back Screen) key.
(15) Press the key of Next Menu to display the first page of main function keys
(F7 to F12).
(16) Press the F9 (Standby Output Power) key.